Seismology Research
Bringing together seismology and atmospheric sciences
4. How can infrasound be measured?
January 2004
Läslo Evers
Measuring infrasound can be achieved via two approaches. Either a microphone can be
made low frequent or a barograph can be made high frequent. At the KNMI both
approaches are applied. In this chapter the high frequent barograph approach is explained, by means
of the KNMI microbarometer and its field application.
4.1 The KNMI microbarometer
In figure 4.1 a schematic view of the KNMI microbarometer is given. Various elements
can be distinguished.
Figure 4.1: A schematic view of the KNMI microbarometer
The atmosphere is probed through the inlet box, shown in the top of figure 4.1. Possible
pressure fluctuations are guided through six inlets towards the small volume inlet box.
The sensing element is a Validyne® pressure transducer, which measures pressure variations
differentially through a variable reluctance scheme. Differential pressure measurement means
the pressure is measured with respect to a reference pressure in a backing volume. This backing
volume is connected beneath the sensor as shown in figure 4.1. Doing so, all possible pressure
fluctuations would be measured also those from meteorological variations. Therefore, a leak is
introduced in the backing volume, being a thin capillary. The acoustical resistance of this
capillary controls the low frequency cut off of the instrument by its diameter and length.
The relaxation time of the instrument is tuned in such a way that the lowest period recovered
equals 500 seconds. The electronics are placed in a box beneath the pressure measuring devices.
Figure 4.2: The KNMI microbarometer
Figure 4.2 shows a picture of the KNMI microbarometer. On top the inlets and inlet box
can be seen. The sensor and backing volume are located within the grey box. The white
box at the bottom contains the electronics. The configuration of the KNMI microbarometer
is such that a robust instrument is obtained, necessary for field applications.
Figure 4.3: The inside of the KNMI microbarometer with backing volume and thin capillary
A view insight the grey box, containing the sensing elements, is given in figure 4.3. The
disc, of stainless steel, is the Validyne® pressure transducer. The backing volume is connected to it.
In figure 4.3, the backing volume is opened to show the thin capillary. Although, the capillary
is difficult to distinguish, it leads pressure fluctuations with periods larger than 500
seconds back to the atmosphere through the white rubber tube connected to the top of the backing
volume. The thick walled backing volume with the capillary inside (instead of outside) garantee
temperature stability.
4.2 Field installation
Measuring pressure differentially introduces temperature dependence; warming up or cooling
down the air within the backing volume will influence the measurements. The instrument is,
therefore, mounted beneath the surface to achieve temperature stability. The
instrument is also protected in this way.
Figure 4.4: Mounting of the KNMI microbarometer in field applications.
In figure 4.4, a schematic view is given of the installation of the KNMI microbarometer
in the field. The instrument is placed in a pvc container, creating a dust, water and
insect proof environment. The latter is experienced non trivial in field applications.
Extra protection of the instrument is achieved by covering it with a concrete top. To each
instrument six porous are connected to reduce wind noise. Wind is coherent over only small
distances, thus by sampling the atmosphere over area rather than one point, noise reduction
is achieved.
Figure 4.5: The underground installation of the KNMI microbarometer at DIA with the
porous hoses connected for wind noise reduction.
Figure 4.5 shows a picture of one of the microbarometer's installations at Airfore base
Deelen in the East of the Netherlands. The intergrating area is defined through six porous
hoses in a spider like layout. To sample the surrounding atmosphere not too close to the
instrument, closed hose (yellow) connects the porous hose (black) to the instrument.